| « Prev |
2025 Apr-Jun; Vol 16, No 2:e6 |
Next » |
|
e6 |
The Utility of Evaluating Primary Stability of Implants Using Accumulated Insertion Torque Values: a Preliminary Study J Oral Maxillofac Res 2025;16(2):e6 doi:10.5037/jomr.2025.16206 Abstract | HTML | PDF |
The Utility of Evaluating Primary Stability of Implants Using Accumulated Insertion Torque Values: a Preliminary Study
1Department of Advanced Prosthodontics, Hiroshima University Graduate School of Biomedical and Health Sciences, Hiroshima, Japan.
Corresponding Author:
Department of Advanced Prosthodontics
Hiroshima University Graduate School of Biomedical Sciences
1-2-3, Kasumi, Minami-ku, Hiroshima 734-8553
Japan
Phone: +81 82 257 5677
Fax: +81 82 257 5679
E-mail: reiko1122@hiroshima-u.ac.jp
ABSTRACT
Objectives: This preliminary in vitro study aimed to clarify the usefulness of the accumulated torque value as a new method for evaluating primary implant stability.
Material and Methods: Implants of different sizes (42-10ST and 37-6ST) and simulated bone blocks of different densities (D1 to D4) were used. The implant osteotomy was prepared following the manufacturer’s drilling protocol in simulated bone blocks, and the implants were inserted (n = 10). The implant insertion torque value was measured at 0.05 seconds intervals. The peak value was recorded as the peak torque value (PTV), and the sum of the torque values measured from the start to the end of the implant insertion was recorded as the accumulated torque value (ATV). After implantation, the implant stability quotient (ISQ) was measured. Correlations among ATV, PTV, and ISQ were analysed.
Results: The ATV increased with the density of the simulated bone block. A strong correlation was observed between the PTV value and ATV value (42-10ST [P < 0.05, r = 0.99], 37-6ST [P < 0.05, r = 0.99]). In addition, a correlation was observed between the ATV and ISQ ISQ (42-10ST [P < 0.05, r = 0.81], 37-6ST [P < 0.05, r = 0.83]). The PTV was higher in the 42-10ST group than in the 37-6ST group from D1 to D4.
Conclusions: The accumulated torque value varied according to the density of the simulated bone block, and a correlation was observed with the existing implant stability evaluation method. This suggests that this method may be useful as a novel approach for evaluating primary implant stability.
J Oral Maxillofac Res 2025;16(2):e6
doi: 10.5037/jomr.2025.16206
Accepted for publication: 30 June 2025
Keywords: bone density; dental implant; dental implantation; torque.
INTRODUCTION
The primary stability of an implant is closely related to treatment prognosis [1,2]. Currently, this primary stability is evaluated using implant stability coefficients, that consider factors such as implant stability quotient (ISQ) and insertion torque (IT) during implant placement [3,4]. The ISQ was measured by attaching a smart peg to the implant, vibrating it with a magnetic pulse, and measuring the resonance frequency of the smart peg. The obtained frequency was converted into the ISQ using a special formula and was expressed as numbers from 1 to 100. It is easy to operate and is used clinically to evaluate implant stability. IT value is measured by the implant micromotor and the evaluation is usually performed by the peak value. In practice, IT is generated over time, and the sum of these torque values is loaded onto the bone surrounding the implant. For this reason, some studies have evaluated the primary stability as the integral value of torque-depth curve or the insertion energy during implant insertion [5-8]. The insertion energy, as evaluated in these reports, is mainly expressed as the integral value of the torque-depth curve and is evaluated as a measure of primary stability. It has been shown to correlate with the immediate bone-to-implant contact. Some surgical motor systems display this torque-depth curve and are occasionally used in clinical practice to evaluate primary stability. However, the implant length, a key parameter in insertion energy, is often evaluated by converting it into the speed of motor rotations. This has the disadvantage that the calculated implant length and actual implant length may differ due to operator variability and differences in bone condition. In this study, torque values were measured at short time intervals, to the nearest 0.1 Ncm, to better account for variables such as implant shape, length, and bone condition. The sum of these IT values, measured at 0.05 second intervals, was used to evaluate the accumulated torque value (ATV).
The American Society for Testing and Materials (ASTM) has approved artificial polyurethane bone blocks as standard material for testing orthopaedic devices and instruments and has declared it an ideal material for comparative testing of bone screws (ASTM F-1839-08) [9]. Corresponding to the different bone densities specified above, the densities of polyurethane bone blocks are also classified into four types [10-12]. According to the Misch bone density classification as D1 = 0.48 g/cc, D2 = 0.32 g/cc, D3 = 0.16 g/cc, and D4 = 0.08 g/cc [10-12]. Several studies have demonstrated that IT and ISQ are correlated with the bone strength or density of the polyurethane bone blocks [5,13-15].
The purpose of this preliminary in vitro study was to determine the usefulness of the accumulated torque value as a novel method for evaluating primary stability by the correlation between the accumulated torque value and the implant stability quotient value.
MATERIAL AND METHODS
Materials
Solid rigid polyurethane-simulated bone blocks of different densities (0.48, 0.32, 0.16, and 0.08 g/cc) and size of 13 cm × 18 cm × 4 cm (SAW 1522-23/01/03/04; Sawbones® - Pacific Research Laboratories, Inc.; Vashon Island, Washington, USA) were used.
Additionally, two different sizes of titanium alloy implant bodies were prepared: 42-10ST (diameter 4.2 mm × length 10 mm, straight-shape, mechanically polished surface - Kyocera Corp.; Kyoto, Japan) and 37-6ST (diameter 3.7 mm × length 6 mm, straight-shape, mechanically polished surface - Kyocera Corp.) (Figure 1). Surgical Pro2 (NSK-Nakanishi International; Kanuma, Tochigi, Japan) was used as the implant motor system.
|
Figure 1 Implants: (a) 42-10ST = diameter 4.2 mm, length 10 mm; (b) 37-6ST = diameter 3.7 mm, length 6 mm. The surfaces were machined and non-treated. |
Methods
The study was conducted at the Department of Advanced Prosthodontics, Hiroshima University Graduate School of Biomedical and Health Sciences, Hiroshima, Japan, between November 13, 2024 to November 22, 2024. An implant osteotomy was prepared in each simulated bone block at 2 cm intervals, and the implants were inserted (n = 10). The implant osteotomies were prepared as follows: round drill, pilot drill, and final drill with a diameter 3.4 mm for the diameter 3.7 implant and diameter 3.9 mm drill (Kyocera Corp.) for the diameter 4.2 implant (1200 rpm, according to the manufacturer’s specified drilling protocol; www.kyocera.co.jp). Implants were inserted at 35 rpm. During insertion, the implant IT value was measured at 0.05 seconds intervals. The data were output from the implant motor system (Surgic Pro2 - NSK-Nakanishi International) to an offline personal computer as an Microsoft Excel file (Microsoft Office Standard 2013 - Microsoft Corp.; Redmond, Washington, USA). The peak value was recorded as the peak torque value (PTV), and the sum of the torque values measured from the start to the end of implant insertion was recorded as the ATV (Figure 2). After implant placement, the ISQ was measured three times along each of the long and short axes, and the average value was recorded 100500 Osstell® ISQ (Osstell AB; Gothenburg, Sweden).
|
Figure 2 The instant torque generated over time was measured every 0.05 seconds. The peak value of these was recorded as the peak torque value (PTV), and the sum of all recorded torque values was recorded as accumulated torque value (ATV). |
Statistical analysis
Pearson’s correlation coefficient (r) was used to evaluate the correlation between ATV-PTV and PTV-ISQ. The ATV, PTV, and ISQ values for each density were compared using t-tests. Data were expressed as mean and standard deviation (M [SD]). Statistical significance level was defined at P < 0.05. The calculations were performed using GraphPad Prism version 7.0 software (GraphPad Software Inc.; La Jolla, CA, USA).
RESULTS
Colour maps of implantation torque values
The implantation torque values were recorded as shown in Figure 3. The higher the density of the bone block, the larger the colour map displayed. Also the larger the implant size, the larger the colour map was displayed. The vertical axis of the graph is the virtual implant length calculated from the rotation speed and time. Although all implants were inserted at 35 rotations, the time required for insertion was not constant, and it was shorter at D4.
|
Figure 3 Colour map of insertion torque: A = D1, 42-10ST; B = D2, 42-10ST; C = D3, 42-10ST; D = D4, 42-10ST; E = D1, 37-6ST; F = D2, 37-6ST; G = D3, 37-6ST; H = D4, 37-6ST. The higher the density of the bone block, the larger the colour map displayed. The larger the implant size, the larger the colour map was displayed. |
In the initial stage of implant placement, the resistance is weak, and the torque value is low, and the resistance gradually increases as the contact area with the bone block increases. The PTV is affected by the block density. The block had a uniform structure, the PTV was recorded when the implant was placed at the deepest depth and the contact area between the block and the implant was at its maximum.
Implants were inserted at 35 rpm, however the time required for insertion was not constant, depending on the bone block density. The implant length derived from the rotational speed and time differed from the actual length of the implanted fixture. In D4 blocks, the insertion process proceeded easily, resulting in a recorded length shorter than the actual depth. Conversely, in D1 blocks with high density and hardness, it had high resistance and, leading to potential rotation without insertion. Thus, implant length derived from rotational speed and insertion time was not precise and may introduce errors in the integration value of torque-depth curves when adjusting insertion depth.
Comparison of measurements between implants
The ISQ of 42-10ST was D1 = 73.9 (1.3), D2 = 65.3 (1.3), D3 = 45.6 (2), D4 = 30.6 (1.5), that of 37-6ST was D1 = 67.5 (1.6), D2 = 54.6 (1.4), D3 = 30.2 (0.7), D4 = 10.6 (2.1), and it was higher in 42-10ST than 37-6ST in each block. The PTV of 42-10ST was D1 = 30.6 (2.6) Ncm, D2 = 9.4 (0.6) Ncm, D3 = 1.5 (0.2) Ncm, D4 = 0.9 (0.1) Ncm, that of 37-6ST was D1 = 26.6 (1.7) Ncm, D2 = 6.9 (0.5) Ncm, D3 = 1.1 (0.1) Ncm, D4 = 0.3 (0.1) Ncm, and it was higher in 42-10ST than 37-6ST in each block. The ATV of 42-10ST was D1 = 4479 (369) Ncm, D2 = 1163 (107) Ncm, D3 = 147.5 (46.7) Ncm, D4 = 76.5 (27.8) Ncm, that of 37-6ST was D1 = 1850 (168) Ncm, D2 = 441.3 (44.5) Ncm, D3 = 54.1 (12.8) Ncm, D4 = 11.7 (11.6) Ncm, and it was higher in 42-10ST than 37-6ST in each block (Table 1).
|
Table 1 Comparison of measurements between implants a-b, c-d, e-f, g-h, i-j, k-l, m-n, o-p, q-r, s-t, u-v, w-xStatistically significant at level P < 0.05 (t-test). SD = standard deviation. |
Correlation
The ATV increased with the density of the simulated bone block. A strong correlation was observed between the PTV and ATV values in both implants (42-10ST [P < 0.05, r = 0.99], 37-6ST [P < 0.05, r = 0.99]) (Figure 4). A correlation was observed between ATV and ISQ (42-10ST [P < 0.05, r = 0.81], 37-6ST [P < 0.05, r = 0.83]) (Figure 5).
|
Figure 4 Correlation between accumulated torque value (ATV) and peak torque value (PTV): A = 42-10ST; B = 37-6ST. Strong correlation was observed in both groups (42-10ST [P < 0.05, r = 0.99], 37-6ST [P < 0.05, r = 0.99]). |
|
Figure 5 Correlation between implant stability quotient (ISQ) and accumulated torque value (ATV): A = 42-10ST; B = 37-6ST. A correlation was observed between the ATV and the ISQ (42-10ST [P < 0.05, r = 0.81], 37-6ST [P < 0.05, r = 0.83]). |
DISCUSSION
This study focused on the IT value as an evaluation of the primary stability of implants and investigated a new evaluation method based on the IT value over time using simulated bone blocks. Insufficient initial fixation may result in a longer healing time or an increased risk of implant disintegration. However, excessively high torque can lead to stress concentration, which may not always produce good outcomes [16-21]. Excessive stress during implant insertion may cause bone resorption, and high IT does not always guarantee positive clinical results [17].
Duyck et al. [22] observed that a high IT of > 45 N increased the risk of bone microfractures, resulting in a greater bone resorption response at the molecular and cellular levels, leading to a significant loss of bone stability within the first stage of healing. Therefore, in implant treatment, it is important to select an appropriate drilling protocol based on bone condition [23].
However, these reports were based on PTV values. The PTV is an evaluation of the value cut-off at a certain moment of the torque usually recorded at the final moment of insertion. Therefore, it did not evaluate all the stresses applied to the bone. The stress applied to the bone during implantation can be comprehensively evaluated by accumulating the IT values applied over time.
In this study, we used simulated bone blocks with a structure like that of cancellous bones. These bone blocks are classified according to Misch classification, and these blocks also showed a significant difference in torque value in our drilling evaluation [14,15]. Implants with different diameters and lengths were examined, implants with larger diameters and lengths exhibited higher ISQ, PTV, and ATV. Furthermore, among the same implants, the higher the block density, the higher the values for all parameters. The PTV and ISQ values were consistent with those in previous reports and correlated well with each other [24,25]. Also, the ATV had a strong correlation between PTV and ISQ. This was because the simulated bone block had a uniform structure.
The ISQ levels of the successfully integrated implants ranged from 57 to 82 [26]. The ISQ value is an evaluation that is influenced by the entire surrounding bone that supports the implant body. A correlation was observed between the ISQ value and the ATV; the higher the integrated torque value, the higher the ISQ value. Therefore, the ATV may also be used to evaluate the entire stress applied around the implant.
For the uniformly structured simulated bone blocks used in this study, a strong correlation was observed between the ATV and PTV. However, in situations where bone density is not uniform, such as in human bones, it might be suggested that evaluation of the ATV may be useful. ATV offers an advantage as it remains unaffected by variations in insertion time or depth changes, allowing for accurate measurement. Currently, as implant bodies with various diameters and lengths are used, the evaluation of ATV may provide a more detailed evaluation of primary implant stability.
Limitations
This study has several limitations:
-
Use of simulated bone blocks may not fully replicate the complex biological responses.
-
Findings may be influenced by the uniform structure of the simulated bone.
-
Limited range of implant sizes and densities may affect the generalizability of the results.
-
Long-term implications of accumulated insertion torque on implant stability and bone health require further investigation in longitudinal studies.
CONCLUSIONS
The accumulated torque value changed according to the density of the simulated bone block for different-sized implants, and a correlation was observed with existing implant stability evaluations. This novel method may be useful for evaluating the stability of primary implants.
ACKNOWLEDGMENTS AND DISCLOSURE STATEMENTS
The authors report no conflicts of interest related to this study.
REFERENCES
- Friberg B, Jemt T, Lekholm U. Early failures in 4,641 consecutively placed Brånemark dental implants: a study from stage 1 surgery to the connection of completed prostheses. Int J Oral Maxillofac Implants. 1991 Summer;6(2):142-6.
[Medline: 1809668] [doi: 10.1097/00008505-199200110-00017] - Javed F, Ahmed HB, Crespi R, Romanos GE. Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation. Interv Med Appl Sci. 2013 Dec;5(4):162-7.
[Medline: 24381734] [PMC free article: 3873594] [doi: 10.1556/imas.5.2013.4.3] - Trisi P, Perfetti G, Baldoni E, Berardi D, Colagiovanni M, Scogna G. Implant micromotion is related to peak insertion torque and bone density. Clin Oral Implants Res. 2009 May;20(5):467-71.
[Medline: 19522976] [doi: 10.1111/j.1600-0501.2008.01679.x] - Degidi M, Daprile G, Piattelli A, Iezzi G. Development of a new implant primary stability parameter: insertion torque revisited. Clin Implant Dent Relat Res. 2013 Oct;15(5):637-44.
[Medline: 22008885] [doi: 10.1111/j.1708-8208.2011.00392.x] - Di Stefano DA, Arosio P, Gastaldi G, Gherlone E. The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks. J Prosthet Dent. 2018 Nov;120(5):706-714.
[Medline: 28689908] [doi: 10.1016/j.prosdent.2017.04.012] - Arosio P, Arosio F, Di Stefano DA. Implant Diameter, Length, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks. Dent J (Basel). 2020 Jun 4;8(2):56.
[Medline: 32512762] [PMC free article: 7345030] [doi: 10.3390/dj8020056] - Grobecker-Karl T, Dickinson A, Heckmann S, Karl M, Steiner C. Evaluation of Insertion Energy as Novel Parameter for Dental Implant Stability. J Clin Med. 2020 Sep 15;9(9):2977.
[Medline: 32942697] [PMC free article: 7565125] [doi: 10.3390/jcm9092977] - Yang B, Irastorza-Landa A, Heuberger P, Ploeg HL. Effect of insertion factors on dental implant insertion torque/energy-experimental results. J Mech Behav Biomed Mater. 2020 Dec;112:103995.
[Medline: 32882675] [doi: 10.1016/j.jmbbm.2020.103995] - Standard specification for rigid polyurethane foam for use as a standard material for testing orthopaedic devices and instruments. ASTM F1839-08(2021). 2021 Feb 10. [doi: 10.1520/F1839-08R21]
- Möhlhenrich SC, Heussen N, Elvers D, Steiner T, Hölzle F, Modabber A. Compensating for poor primary implant stability in different bone densities by varying implant geometry: a laboratory study. Int J Oral Maxillofac Surg.2015 Dec;44(12):1514-20.
[Medline: 26362488] [doi: 10.1016/j.ijom.2015.08.985] - Möhlhenrich SC, Kniha K, Heussen N, Hölzle F, Modabber A. Effects on primary stability of three different techniques for implant site preparation in synthetic bone models of different densities. Br J Oral Maxillofac Surg.2016 Nov;54(9):980-986.
[Medline: 27461557] [doi: 10.1016/j.bjoms.2016.07.004] - Comuzzi L, Iezzi G, Piattelli A, Tumedei M. An In Vitro Evaluation, on Polyurethane Foam Sheets, of the Insertion Torque (IT) Values, Pull-Out Torque Values, and Resonance Frequency Analysis (RFA) of NanoShort Dental Implants. Polymers (Basel). 2019 Jun 10;11(6):1020.
[Medline: 31185590] [PMC free article: 6630510] [doi: 10.3390/polym11061020] - Di Stefano DA, Piattelli A, Iezzi G, Orlando F, Arosio P. Cortical Thickness, Bone Density, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks. Int J Oral Maxillofac Implants. 2021 May-Jun;36(3):423-431.
[Medline: 34115054] [doi: 10.11607/jomi.8419] - Wakamatsu K, Doi K, Kobatake R, Makihara Y, Oki Y, Tsuga K. Investigation to Predict Primary Implant Stability Using Frictional Resistance Torque of Tap Drilling. J Oral Maxillofac Res. 2022 Dec 31;13(4):e1.
[Medline: 36788798] [PMC free article: 9902023] [doi: 10.5037/jomr.2022.13401] - Wakamatsu K, Doi K, Kobatake R, Oki Y, Tsuga K. Evaluation of Bone Density for Primary Implant Stability Using a Newly Designed Drill: An In Vitro Study on Polyurethane Bone Blocks. Clin Exp Dent Res. 2024 Dec;10(6):e70048.
[Medline: 39610027] [PMC free article: 11604594] [doi: 10.1002/cre2.70048] - Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent. 1983 Jul;50(1):101-7.
[Medline: 6576145] [doi: 10.1016/0022-3913(83)90174-9] - Bashutski JD, D'Silva NJ, Wang HL. Implant compression necrosis: current understanding and case report.J Periodontol. 2009 Apr;80(4):700-4.
[Medline: 19335092] [doi: 10.1902/jop.2009.080581] - Campos FE, Gomes JB, Marin C, Teixeira HS, Suzuki M, Witek L, Zanetta-Barbosa D, Coelho PG. Effect of drilling dimension on implant placement torque and early osseointegration stages: an experimental study in dogs.J Oral Maxillofac Surg. 2012 Jan;70(1):e43-50.
[Medline: 22182660] [doi: 10.1016/j.joms.2011.08.006] - Menicucci G, Pachie E, Lorenzetti M, Migliaretti G, Carossa S. Comparison of primary stability of straight-walled and tapered implants using an insertion torque device. Int J Prosthodont. 2012 Sep-Oct;25(5):465-71.
[Medline: 22930768] - Jimbo R, Tovar N, Anchieta RB, Machado LS, Marin C, Teixeira HS, Coelho PG. The combined effects of undersized drilling and implant macrogeometry on bone healing around dental implants: an experimental study. Int J Oral Maxillofac Surg. 2014 Oct;43(10):1269-75.
[Medline: 24794761] [doi: 10.1016/j.ijom.2014.03.017] - Tabassum A, Meijer GJ, Walboomers XF, Jansen JA. Evaluation of primary and secondary stability of titanium implants using different surgical techniques. Clin Oral Implants Res. 2014 Apr;25(4):487-92.
[Medline: 23638908] [doi: 10.1111/clr.12180] - Duyck J, Corpas L, Vermeiren S, Ogawa T, Quirynen M, Vandamme K, Jacobs R, Naert I. Histological, histomorphometrical, and radiological evaluation of an experimental implant design with a high insertion torque. Clin Oral Implants Res.2010 Aug;21(8):877-84.
[Medline: 20528892] [doi: 10.1111/j.1600-0501.2010.01895.x] - Fernández-Olavarria A, Gutiérrez-Corrales A, González-Martín M, Torres-Lagares D, Torres-Carranza E, Serrera-Figallo M. Influence of different drilling protocols and bone density on the insertion torque of dental implants. Med Oral Patol Oral Cir Bucal. 2023 Jul 1;28(4):e385-e394.
[Medline: 37330951] [PMC free article: 10314359] [doi: 10.4317/medoral.25804] - O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res. 2000;2(2):85-92.
[Medline: 11359268] [doi: 10.1111/j.1708-8208.2000.tb00110.x] - do Vale Souza JP, de Moraes Melo Neto CL, Piacenza LT, Freitas da Silva EV, de Melo Moreno AL, Penitente PA, Brunetto JL, Dos Santos DM, Goiato MC. Relation Between Insertion Torque and Implant Stability Quotient: A Clinical Study. Eur J Dent. 2021 Oct;15(4):618-623.
[Medline: 34233364] [PMC free article: 8630976] [doi: 10.1055/s-0041-1725575] - Balleri P, Cozzolino A, Ghelli L, Momicchioli G, Varriale A. Stability measurements of osseointegrated implants using Osstell in partially edentulous jaws after 1 year of loading: a pilot study. Clin Implant Dent Relat Res. 2002;4(3):128-32.
[Medline: 12516644] [doi: 10.1111/j.1708-8208.2002.tb00162.x]
|
To cite this article: The Utility of Evaluating Primary Stability of Implants Using Accumulated Insertion Torque Values: a Preliminary Study J Oral Maxillofac Res 2025;16(2):e6 URL: http://www.ejomr.org/JOMR/archives/2025/2/e6/v16n2e6ht.htm |
Received: 26 June 2025 | Accepted: 30 June 2025 | Published: 30 June 2025
Copyright: © The Author(s). Published by JOMR under CC BY-NC-ND 3.0 licence, 2025.







.jpg)
.jpg)
.jpg)
.png)
.jpg)
.jpg)